power and energy : ultrapure palladium diffusion hydrogen purifiers



 Return to Products List


  Download the Spec Sheet for the 6000/7000:
  Power+Energy 6000/7000 Series Specs (696KB PDF)




 
6000/7000 Series High-Flow Hydrogen Purifier Systems

Introduction:

Power+Energy's 6000- and 7000-Series hydrogen purifiers are designed for use as cost-effective new or replacement cells for use in your existing equipment. Hydrogen purity levels of 99.9999999% are achievable with continuous operation at flow rates ranging from 60 to 400 standard liters per minute, whether supplied by commercial hydrogen cylinders or by an in-house hydrogen supply.

Whether purchased or rented as an Operational Lease, each hydrogen purifier is fully helium-leak-tested during assembly and bench-tested before shipping in order to ensure the longest lifetime possible.

Principle of Operation:

Power+Energy hydrogen purifiers consist of a 316L-grade stainless steel cell, incorporating a large-diameter palladium-alloy tube as the diffusion element. Hydrogen gas is brought into contact with the inner surfaces of the palladium-alloy tube, which acts as a selective barrier, passing only atomic hydrogen through the tube wall while excluding all other gases.

Molecular hydrogen is adsorbed onto the surface, where it dissociates into atomic hydrogen. Each hydrogen atom then shares its electron with the palladium-alloy lattice, and diffuses through the lattice as determined by the pressure gradient. The hydrogen atoms recombine on the low-pressure side of the membrane, and are desorbed as hydrogen molecules with virtually no impurities.

The principal benefit of Power+Energy's inside-out architecture, compared to competitive models' outside-in methods, is that Power+Energy provides far cleaner hydrogen with far less maintenance — the outside surface of the large-diameter tubing can easily be thoroughly cleaned, eliminating dead spots and increasing both efficiency and purity.

Applications:

Power+Energy's hydrogen purifiers are ideal for use in MOCVD reactors, laboratories, universities, or anywhere else ultra-pure hydrogen is needed. Popular applications include:

•  Semiconductor manufacturing

•  Use as a carrier gas and fuel in gas chromatography

•  Metallurgy

•  Hydrogenation

•  Catalytic processes

•  Controlled atmospheres

•  Gas treatment in nuclear power plants

Locally generated ultra-pure hydrogen is ideal for use as a carrier gas and for flame ionization in gas chromatography applications.

Product Highlights:
•  Highly reliable and efficient P+E inside-feed palladium cell design
•  Compact design for large UPH flows
•  Proportional, differential, integrating temperature controller
•  Accelerated Life Testing ensures reliability
•  One-year warranty

Mechanical Specifications:

•  Tubing/Cell Material:                           316L stainless steel
•  Helium Leak Test:                               1E-9 cc/sec
•  Back Panel Tubing Connections:        1/4" face seal fittings
•  Membranes Tested to 400 PSIG

•  Finish on all UPH tubing electro-polished
•  Finish on all UPH cell surfaces micro-polished

Process Specifications:

•  Ultra-Pure Hydrogen Purity:                 99.9999999%

      (at UPH flow > 20% of max flow or > 15 SLPM, whichever is less)
•  Ultra-Pure Hydrogen Flow Rate:          60-400 SLPM (model dependent)
•  Hydrogen Gas Feed Inlet Pressure:     36 - 200 PSIG (model dependent)

•  Feed Hydrogen Gas Purity:                  ~99.95%
•  Hydrogen Bleed Outlet:                         typically 1 to 4% of feed hydrogen
•  Typical Average Power Consumption: <700 watts

Sizing:
The 6000 and 7000 Series purifiers are available in a number of different models in order to accomodate your gas output flow and spacing requirements. The flow rates across the 6000/7000 series range from 60 to 400 standard liters per minute (SLPM). Please note that the physical dimensions will vary by model, depending on maximum output.

  

Model

  

Output (SLPM)

Length
(Inches)
Height
(Inches)
  Model 6060   60   15.6   13.7
  Model 6080   80   20.4   13.7
  Model 6100   100   20.4   13.7
  Model 6130   130   25.2   13.7
  Model 6160   160   30.0   13.7
  Model 7080   80   9.7   14.8
  Model 7145   145   14.6   14.8
  Model 7160   160   14.6   14.8
  Model 7180   180   19.5   14.8
  Model 7195   195   19.5   14.8
  Model 7210   210   19.5   14.8
  Model 7220   220   19.5   14.8
  Model 7245   245   19.5   14.8
  Model 7260   260   24.4   14.8
  Model 7280   280   24.4   14.8
  Model 7290   290   24.4   14.8
  Model 7310   310   24.4   14.8
  Model 7325   325   24.4   14.8
  Model 7340   340   29.3   14.8
  Model 7360   360   29.3.   14.8
  Model 7375   375   29.3   14.8
  Model 7390   390   29.3   14.8
  Model 7405   400   29.3   14.8

Flow Curve:

(Model 7150 flow curve shown. Others available on request.)


(Click for full size)



 
     
Copyright © 2008 Power+Energy, Inc.